New Hybrid Xcela PICC

Is anyone using the new Hybrid Xcela PICC with dedicated line for CVP monitoring? Wondering if it is accurate with CVP readings etc and overall thoughts about the product.  Also interested in any studies out there on it if anyone can direct me to them.  Thanks

lynncrni
Use of a PICC for CVP

Use of a PICC for CVP monitoring has been documented. This hybrid PICC means that one lumen does not have the pressure sensitive valve in the external catheter hub, while all other lumens do have this valve. There should be no problem with measuring accurate CVPs on the lumen without the valve, but only on that specific lumen. Lynn

Lynn Hadaway, M.Ed., RN, BC, CRNI

Lynn Hadaway Associates, Inc.

126 Main Street, PO Box 10

Milner, GA 30257

Website http://www.hadawayassociates.com

Office Phone 770-358-7861

lucy jones
I have used the Xcela hybrid

I have used the Xcela hybrid and LOVED it!  Wish I still had access to it. 

 